| A conventional cardiac
catheterization system uses a point x-ray source
to project an x-ray beam through the patient onto
a large-area detector. NovaRay’s system
has an inverse geometry, which consists of a large-area
scanning x-ray source and a small-area detector.
The scanning x-ray source projects a rapid sequence
of narrow x-ray beams through the patient and
onto a small-area, high efficiency detector. The
system acquires a series of overlapping small-field-of-view
images that are combined electronically to generate
the full image.
A major advantage of inverse geometry is its inherent
scatter reduction. A conventional system has a
large-area detector close to the patient’s
chest and, therefore, a large amount of scatter
radiation reaches the detector. To address this
problem, a conventional system incorporates a
grid—a parallel arrangement of thin lead-foil
strips—between the patient and the detector,
to reduce the amount of scatter radiation reaching
the detector. Due to the thickness of the lead-foil
strips, approximately 30% of the image radiation
is absorbed in the scatter-reduction grid, reducing
image quality. Because this grid is an imperfect
device, typically as much scatter radiation as
image radiation reaches the detector. The scatter
radiation is noise that lowers the signal-to-noise
ratio and reduces image quality.
NovaRay’s system has a small-area detector
at a large distance from the patient. With this
geometry, very few of the x-ray photons scattered
randomly off the patient strike the detector.
With much less scattered radiation, there is a
significant reduction in the noise source that
reduces image quality in a conventional system.
By eliminating the need for a scatter-reduction
grid, NovaRay also avoids the associated loss
of image radiation.
The NovaRay inverse geometry also spreads the
incident radiation over a larger area of the patient’s
skin. The point x-ray source of a conventional
system concentrates the radiation where it enters
the patient’s skin. With NovaRay’s
distributed x-ray source, x-ray radiation is diffused
where it enters the patient’s skin, spreading
radiation over approximately double the area of
a conventional system, cutting skin radiation
exposure by 50% and reducing the possibility of
radiation injury.
Inverse geometry also allows for open patient
access. The large-area detector in a conventional
system must be kept close to the patient’s
chest for optimum image quality. This arrangement
is claustrophobic for the patient and restricts
access of the clinical staff to the patient. In
contrast, the NovaRay detector is approximately
one meter away from the patient, allowing easy
patient access at all times.
|